Abstract

We encountered a case in which prosthetic mitral valve leaflets were jammed by a transmitral balloon catheter during mechanical mitral valve replacement. The catheter, which was inserted in the left ventricle to render the leaflets incompetent during de-airing, forced the leaflets into the closed position. The leaflet edges pinched the catheter lumen and obstructed balloon deflation. Excessive inflation of the balloon and extraction of the catheter before complete deflation of the balloon should be avoided to prevent a potentially serious complication.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.