Abstract

We aimed to assess the prevalence and features of prosthesis-patient mismatch (PPM) following transcatheter aortic valve implantation (TAVI) and its prognostic impact considering baseline left ventricular ejection fraction (LVEF). Data from 1,309 patients undergoing TAVI for severe aortic stenosis were derived from a single-centre dedicated TAVI registry. PPM was assessed according to echocardiography at discharge and was defined in accordance with VARC-2. Median follow-up time was 2.03 years. Moderate and severe PPM was detected in 22.9% and 12.9%, respectively. Patients with severe PPM had smaller annuli and more often received transcatheter heart valve (THV) sizes ≤23 mm. Supra-annular THV design showed the lowest rate of PPM. In patients with LVEF <40%, but not in those with LVEF ≥40%, severe PPM was associated with an increased three-year mortality rate (no vs. severe PPM for LVEF ≥40%: 34.6% vs. 29.5%, p=0.96; LVEF <40%: 45.1% vs. 68.0%, p=0.041) and was independently predictive of all-cause mortality according to multivariate analysis in these patients (HR 2.97; 95% CI: 1.58-5.59, p<0.001). The presence of severe PPM depends on annular dimensions and THV size and design and is an independent predictor of mortality in patients with reduced LVEF. Hence, the risk of PPM should be considered within the process of THV selection.