Abstract

BackgroundCemented stems show good long-term results and the survival of new implants can be predicted by their early subsidence. With EBRA-FCA (Femoral Component Analysis using Einzel-Bild-Röntgen-Analyse) early subsidence as an early indicator for later aseptic loosening can be analysed. For the cemented TwinSys stem mid- and long-term data is only avalible from the New Zeeland Arthroplasty register, thus close monitoring of this implant system is still mandatory.MethodsWe conducted a 2 year follow up of 100 consecutive hybrid THA (Total hip arthroplasty) of a series of 285 primary THA operated between Jan 2009 und Oct 2010. These 100 received a polished, cemented collarless straight stem (twinSys®, Mathys AG® Bettlach, Switzerland) with an uncemented monobloc pressfit cup (RM pressfit®, Mathys AG® Bettlach, Switzerland). The other patients were treated with the uncemented version of this stem and the same cup. Clinical (Harris Hip Score) and radiological (ap and axial x-rays, cementing quality according to Barrack, alignment) outcomes besides an EBRA-FCA subsidence analysis were performed.ResultsMedian age at operation was 78 (68 to 93) years. 5 patients died in the course of follow-up unrelated to surgery. The KM (Kaplan-Meier) survival at 2 years for the endpoint reoperation for any reason was 94.9 (95 % confidence interval 90.6–100 %). Survival for the endpoint aseptic loosening at 2 years was 100 %.The HHS (Harris Hip Score) improved from 56 (14–86) preoperatively to 95 (60–100) 2 years after the operation. Cementing results were judged 47 % Grade A, 45 % Grade B and 7 % Grade C.Osteolysis was found in 2 stems without clinical symptoms or correlation to subsidence or cementing quality. The EBRA-FCA analysis showed an average subsidence of -0.30 mm (95 % CI -0.5 mm to -0.1 mm). 11 patients showed a subsidence of more than 1 mm. In this group one patient showed a subsidence of 1.5 mm and one of 3.1 mm without further radiological changes.ConclusionsThe twinSys stem showed excellent clinical and radiologic short term results at 2 years follow-up and seems to be a reliable implant.

Highlights

  • Cemented stems show good long-term results and the survival of new implants can be predicted by their early subsidence

  • Different cut-off values, depending on the means of measurement, are described [1,2,3,4]. These different measurements can be performed with plain radiographs, EBRA-FCA (Femoral Component Analysis using Einzel-Bild-Röntgen-Analyse) or Roentgen stereophotogrammetric analysis (RSA) (Roentgen Stereophotogrammetric Analysis)

  • Between Jan 2009 and Oct 2010 a total of 285 primary total hip arthroplasty (THA) were performed at our institution, 100 (97 patients) of them were operated with the cemented twinSys straight stem and followed prospectively after 6, 12, 52 and 104 weeks

Read more

Summary

Introduction

Cemented stems show good long-term results and the survival of new implants can be predicted by their early subsidence. With EBRA-FCA (Femoral Component Analysis using Einzel-Bild-Röntgen-Analyse) early subsidence as an early indicator for later aseptic loosening can be analysed. Long-term success of a cemented stem depends on the longevity of the cement-bone and the cement-prosthesis interface. Different cut-off values, depending on the means of measurement, are described [1,2,3,4] These different measurements can be performed with plain radiographs, EBRA-FCA (Femoral Component Analysis using Einzel-Bild-Röntgen-Analyse) or RSA (Roentgen Stereophotogrammetric Analysis). The Exeter stem is one of the most successful stem designs with outstanding long-term results [8,9,10,11,12,13]

Methods
Results
Discussion
Conclusion