Abstract

Introduction : Appropriate size selection of ProSeal laryngeal mask airway (PLMA) is an important prerequisite for successful use of the device. Size of PLMA is often selected based on weight of patients. We aimed to determine whether combined width of patient’s index, middle and ring fingers could be used as an alternative to weight based selection of appropriate size of PLMA. Materials and methods : In this prospective randomized study, 102 patients aged 6 months to 10 years were included. Patients were allocated to one of two groups based on a randomization table and PLMA inserted according to the group allocated. In Group S, PLMA size selected according to weight based method whereas in Group N, PLMA size was selected based on 3-finger breadth of patients. We recorded insertion parameters, fiberoptic scoring of glottic visualisation, airway manipulations and postoperative complications. Results : In 68 children (66.66%) size of PLMA was same according to both methods. The kappa coefficient was 0.461 concluding moderate agreement between two methods. In the remaining 34 children (33.34%), the disagreement in sizes were with only one size, in whom the patient’s weights were borderline values. The PLMA insertion time, ease of insertion, insertion attempts, fiberoptic view of position of device were comparable in both groups. There were no significant complications in the postoperative period. Conclusion : Three finger breadth sizing method can be used as a good alternative to weight-based method for selecting appropriate size of PLMA in children.

Highlights

  • Appropriate size selection of ProSeal laryngeal mask airway (PLMA) is an important prerequisite for successful use of the device

  • Ventral surface of cuff of PLMA was applied against palmar side of index, middle and ring fingers held in approximation

  • In 68 patients (66.66%), the size of the PLMA coincided with both the methods, with calculated kappa coefficient being 0.461

Read more

Summary

Introduction

Appropriate size selection of ProSeal laryngeal mask airway (PLMA) is an important prerequisite for successful use of the device. Size of PLMA is often selected based on weight of patients. We aimed to determine whether combined width of patient’s index, middle and ring fingers could be used as an alternative to weight-based selection of appropriate size of PLMA. Patients were allocated to one of two groups based on a randomization table and PLMA inserted according to the group allocated. In Group S, PLMA size selected according to weight-based method whereas in Group N, PLMA size was selected based on 3-finger breadth of patients. Conclusion: Three finger breadth sizing method can be used as a good alternative to weightbased method for selecting appropriate size of PLMA in children. Owing to the anatomical variations of upper airway as well as wider range of body weight as a selection criterion for a given size of PLMA, the weight-based size selection recommended by manufacture may not suit all patients.[4,5]

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.