Abstract

PurposeTo study the use of the self-expanding Covera covered stent for the treatment of stenotic lesions at the venous anastomosis of hemodialysis arteriovenous grafts (AVGs). Materials and MethodsA total of 110 patients with AVG venous anastomotic stenosis of ≥50% and access dysfunction were treated at 14 centers in the United States using percutaneous transluminal angioplasty followed by covered stent placement. The primary end points were 30-day safety and 6-month target lesion primary patency (TLPP). The secondary measures included access circuit primary patency, circuit cumulative patency, and the number of reinterventions through 24 months. ResultsFreedom from a primary safety event at 30 days was 96.4% while the 6-month TLPP rate was 70.3%. Seventy-five patients completed 24-month follow-up (68.2%). The TLPP rates were 54.2% at 365 days and 36.9% at 730 days while the access circuit primary patency rates were 16.7% at 365 days and 7.8% at 730 days (Kaplan-Meier analysis). The access circuit cumulative patency rates were 85.4% at 12 months and 73.6% at 24 months. The number of reinterventions to maintain the patency of the access circuit was 3.6 ± 3.1 at 24 months (1.6 ± 1.9 at the target lesion). ConclusionsUse of the Covera covered stent for hemodialysis graft-vein anastomotic stenosis provided a safe treatment option with a TLPP rate of 70.3% at 6 months and TLPP and cumulative access circuit patency rates of 36.9% and 73.6% at 2 years.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call