Abstract
Hypoglycemia, if recurrent, may have severe consequences on cognitive and psychomotor development of neonates. Therefore, screening for hypoglycemia is a daily routine in every facility taking care of newborn infants. Point-of-care-testing (POCT) devices are interesting for neonatal use, as their handling is easy, measurements can be performed at bedside, demanded blood volume is small and results are readily available. However, such whole blood measurements are challenged by a wide variation of hematocrit in neonates and a spectrum of normal glucose concentration at the lower end of the test range. We conducted a prospective trial to check precision and accuracy of the best suitable POCT device for neonatal use from three leading companies in Europe. Of the three devices tested (Precision Xceed, Abbott; Elite XL, Bayer; Aviva Nano, Roche), Aviva Nano exhibited the best precision. None completely fulfilled the ISO-accuracy-criteria 15197: 2003 or 2011. Aviva Nano fulfilled these criteria in 92% of cases while the others were <87%. Precision Xceed reached the 95% limit of the 2003 ISO-criteria for values ≤4.2mmol/L, but not for the higher range (71%).Although validated for adults, new POCT devices need to be specifically evaluated on newborn infants before adopting their routine use in neonatology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.