Abstract

Values for diagnostic reference levels are proposed for intraoral dental radiography in the Republic of Serbia. Proposed numbers for diagnostic reference levels were based on measurements on 119 intraoral units in clinical settings for an adult posterior bitewing X-ray. Values of 3.17 mGy, 3.11 mGy, and 1.58 mGy incident air kerma were found for image receptors film E class, charge-coupled device, and photostimulable phosphor plate, respectively. Similarly, 89.6 mGycm2, 88 mGycm2, and 44.6 mGycm2 air kerma-area products were found for the same detectors. All values are greater than those published in several other similar studies. This is especially true for charge-coupled device image receptors since this technology enables lower patient doses as reported before. This calls for urgent action to optimize intraoral dental imaging, so we present the radiographic techniques used and equipment description to help guide optimization actions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.