Abstract

The reliability of forensic inorganic gunshot residue (iGSR) analysis depends strongly on the performance of the scanning electron microscope with energy dispersive X-ray spectrometer (SEM/EDS) instrument and the software that automates the data acquisition. Without a record of the baseline performance of the instrument, it is often hard to know when performance has deteriorated. This paper provides a set of tests which can be performed when the instrument is commissioned, or when significant modifications are made, to establish the baseline performance and a smaller set of tests that can be performed regularly to document the current performance of the instrument. Together, these represent a powerful way to establish, document and evaluate the performance of the instrumentation used for iGSR analysis. Furthermore, most of these tests are suitable for characterizing SEM/EDS instruments used for other forensic or non-forensic uses.