Abstract

A method for circular e−/e+ accelerator beam energy measurement is proposed. A coordinate of an electron (or a positron) in a focusing magnetic spectrometer built in a circular accelerator depends on its energy, two spectrometer parameters, and the circulating beam energy. The spectrometer parameters can be determined using minimum electron energies from Compton backscattering with two laser wavelengths, and a coordinate detector for photons. The photon detector is calibrated separately at well-known beam energy. The VEPP-4M collider has appropriate equipment for the method implementation: built-in focusing magnetic spectrometer, Compton calibration system with two lasers and a photon coordinate detector. Thus, the proposed technique could be implemented with minimum efforts; tests and further upgrade are planned. The beam energy can be defined with expected uncertainty better than 10−4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.