Abstract

AbstractA flow proportional counter for ultrasoft x-ray applications has been constructed and used to measure proportional counter parameters in the ultrasoft x-ray region at atmospheric and subatmospheric pressures for counting gases of P-10 and methane. We find that at atmospheric pressures the parametrization given by Zastawny for describing gas amplification in proportional counters applies and have measured values for the parameters for P-10 and methane. For subatmospheric pressures the Zastawny parameters are inadequate. A new parametrizatlon is required. In addition, the linearity of the detector for x-rays of energies of 15 KeV to 0.5 KeV was measured for pressures ranging from 750 Torr to 150 Torr and found to be excellent in all pressure regions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.