Abstract

The advent of new arthroscopic devices has led to the development of novel techniques of arthroscopic rotator cuff repair. However, failure to recognize specific technical aspects and improper application of these devices can lead to complications. We report a case of intra-articular protrusion of knotless anchors (PEEK PushLock SP, 4.5 × 18.5 mm; Arthrex, Naples, FL), used in the lateral row of a suture-bridging technique for arthroscopic rotator cuff repair. This case draws attention to the increased length of such devices when compared with traditional suture anchors, the technical aspects of proper device use, the possible patient-related factors such as implant-patient size mismatch, and the importance of additional imaging for the investigation of failure to progress postoperatively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.