Abstract

In this article, we describe our experience with shear wave propagation imaging (SWPI) as an adjunct to 2-dimensional (2D) shear wave elastography (SWE) in a cohort of patients being evaluated for diffuse liver disease. Two-dimensional SWE has been extensively studied in previous publications; however, 2D SWE using propagation images has not been widely described in the literature to date. We observed that when certain artifacts occurred on the color elastograms, highly characteristic changes to shear wave propagation contours were seen, which can help clarify the cause of the artifacts. To our knowledge, the use of SWPI to explain the etiology of artifacts has never been published before. The artifacts described in this article include the capsule reverberation artifact, penetration limitation or dropout artifact, artifact due to blood vessels, shadowing artifact, tissue motion artifact, and near-field distortion/precompression artifact. Hence, the purpose of this article is to show examples of common artifacts seen on 2D SWE as depicted on corresponding SWPI to demonstrate that both types of image displays are complementary to each other.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.