Abstract

Prompt-gamma neutron activation (PGNA) analysis is used for the non-invasive measurement of human body composition. Advancements in portable, compact neutron generator design have made those devices attractive as neutron sources. Two distinct generators are available: D-D with 2.5 MeV and D-T with 14.2 MeV neutrons. To compare the performance of these two units in our present PGNA system, we performed Monte Carlo simulations (MCNP-5; Los Alamos National Laboratory) evaluating the nitrogen reactions produced in tissue-equivalent phantoms and the effects of background interference on the gamma-detectors. Monte Carlo response curves showed increased gamma production per unit dose when using the D-D generator, suggesting that it is the more suitable choice for smaller sized subjects. The increased penetration by higher energy neutrons produced by the D-T generator supports its utility when examining larger, especially obese, subjects. A clinical PGNA analysis design incorporating both neutron generator options may be the best choice for a system required to measure a wide range of subject phenotypes.