Abstract

A focusing neutron lens using glass polycapillary fibers has been introduced successfully into a prompt gamma activation analysis (PGAA) instrument placed at the exit of a cold neutron guide. The neutron current density gain of the lens is 80, averaged over the focused beam size of 0.53 mm diameter. PGAA measurements have been made on submillimeter particles of gadolinium and cadmium. The results indicate that elemental sensitivities of measurements are increased by ∼ 60, and that particles of sizes smaller than 0.5 mm can be discerned using the focusing lens. The measured gain in prompt gamma signals for these particles is less than anticipated, probably due to alignment difficulties. Gamma ray background associated with the lens is discussed and improvements are suggested.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.