Abstract
This study highlights that the currently employed analytical approach for the ICP-OES determination of 233U at λ = 385.950 nm is neither specific for this U isotope nor reliable and provides potential solutions for a selective and reliable quantification of 233U, 235U, and 238U using high resolution ICP-OES. While being a sensitive U wavelength region, the small isotopic line splitting of 233U (λ = 385.950 nm), 235U (λ = 385.955 nm) and 238U (λ = 385.958 nm) cannot be resolved with any commercial high resolution optical spectrometer. In addition, the width of 5.67 pm of the 233U peak, defined as full width at half maximum, is particularly larger than the 238U peak width of ∼4.3 pm, thus indicating splitting of the 233U emission signal into multiple peaks. In contrast, superior U isotopic information can be obtained from isotope specific emission lines centred around λ = 411.6 nm and λ = 424.4 nm with a spectral distance between 233U and 238U of 25 pm and 39 pm, respectively. The 233U signal of both wavelength regions, however, suffers from a potential interference with thorium that requires peak deconvolution for accurate quantification of 233U.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.