Abstract

To evaluate the time to occlusion alarm for peristaltic infusion devices used in Toronto adult critical care units. Cross-sectional study. Biomedical engineering departments of four Toronto teaching hospitals. Twenty peristaltic infusion devices (five Sigma 8000-plus, five Graseby 3000, five Baxter Colleague, and five Alaris 7230B). None. Time to occlusion alarm at flow rates of 2, 10, and 100 mL/hr at a full range of available pressure thresholds for occlusion detection, and with commonly used tubing sets. At default (mid-range) pressure thresholds, mean (SD) time to occlusion alarm was 0.3 (0.1) min at a flow rate of 100 mL/hr, 2.3 (0.5) min at a flow rate of 10 mL/hr, and 11.7 (3.1) min at a flow rate of 2 mL/hr. Time to occlusion alarm in peristaltic infusion devices is long at low flow rates. Patients receiving important medications with short half-lives at low flow rates could experience clinically important interruptions in treatment. Time to occlusion alarm at high flow rates is short, which could lead to excessive alarms and "alarm mistrust" by clinical staff.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.