Recent experiments with the National Institute of Standards and Technology (NIST) Electrostatic Force Balance (EFB) have achieved agreement between an electrostatic force and a gravitational force of 10−5 N to within a few hundred pN/μN. This result suggests that a force derived from measurements of length, capacitance, and voltage provides a viable small force standard consistent with the Système International d’Unités. In this paper, we have measured the force sensitivity of a piezoresistive microcantilever by directly probing the NIST EFB. These measurements were linear and repeatable at a relative standard uncertainty of 0.8%. We then used the calibrated cantilever as a secondary force standard to transfer the unit of force to an optical lever–based sensor mounted in an atomic force microscope. This experiment was perhaps the first ever force calibration of an atomic force microscope to preserve an unbroken traceability chain to appropriate national standards. We estimate the relative standard uncertainty of the force sensitivity at 5%, but caution that a simple model of the contact mechanics suggests errors may arise due to friction.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.