Abstract

Following exposure to room ambient air, the radiation response of orthogonal strip detectors fabricated from high purity germanium for use in a gamma camera has been found to undergo deterioration due to changes which take place at the exposed surfaces between the electrode strips and at the edges of each device. Based upon data obtained from small test detectors subjected to a commercial passivation process, long term stability (greater than one year) following room ambient exposure appears feasible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.