Abstract

Extracorporeal CO2 removal combined with low-frequency positive pressure ventilation (ECCO2-R LFPPV) is a new therapeutic approach in treatment of ARDS. The main problem during long-term extracorporeal support is anticoagulation and related bleeding problems. We conducted a prospective, randomized and controlled clinical trial in 18 patients to compare the effect of the non-heparin-coated (Scimed = group 1) with the heparin-coated (Carmeda = group 2) extracorporeal circuit on clinical course and complication rate. In group 2 the daily blood loss, the amount of substituted red cells and the i.v. heparin dose were significantly lower than in group 1. Bleeding complications were less and more patients survived in group 2. The disadvantage of the hollow fiber oxygenators in the heparin-coated system was plasma leakage, which was more frequent in patients with pancreatitis and hyperbilirubinemia.