Abstract

Two color laser diagnostics using THz laser sources are under development for a high performance operation of the Large Helical Device and for future fusion devices such as ITER. So far, we have achieved high power laser oscillation lines simultaneously oscillating at 57.2 μm (1.6 W) and 47.7 μm (0.8 W) by using a twin optically-pumped CH3OD laser. And, we confirmed the usefulness of this two color laser operation for two-color laser interferometry in a test stand. For the measurement of the Faraday rotation angle we have developed a photo-elastic modulator (PEM) operating around 50 μm, and performed bench tests of the polarimeter with dual PEMs and demonstrated the feasibility of the polarimeter. The achieved angular resolution is 0.01 degrees with a time resolution of 1 ms, which satisfies diagnostic requirements for q-profile measurements on ITER (0.05 deg. at 10 ms time resolution). In this symposium, recent hardware developments and simultaneous measurements of a phase shift and a rotation angle will be presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.