Abstract

The aim of this work was to develop a programmable flow system for rapid assessment of total antioxidant capacity (TAC). Novel features of the prototype include a single pressure-driven source, versatile manipulation of fluid flows, ability to adapt to different TAC assays, and compatibility with microfluidic design. Antioxidant activity was determined by electrochemical measurement of residual 2,2-diphenyl-2-picrylhydrayl hydrate (DPPH•) free radicals in the solution. The overall performance of the device was validated by the spectrophotometric method. The results indicated that dosing of reagents and samples could be controlled by pressure (R2 = 0.992) and time (R2 = 0.999) with high accuracy, and the mixing uniformity of the device was equivalent to that of the batch mixing (R2 = 0.994). TAC assays of a standard antioxidant, ascorbic acid, as well as selected samples such as orange and pomegranate juices, white wine, and green tea by the device were comparable to traditional measurement techniques. Due to the short incubation time, the approach may be more suitable for fast, rather than slow reacting antioxidant compounds. The developed system could be used for rapid TAC screening of food and biological samples.