Abstract

Despite a few studies evaluating the prognostic impact of coronary chronic total occlusion (CTO) in implantable cardioverter defibrillator (ICD) recipients, the impact of CTO on different types of recurrences of ventricular tachyarrhythmias, as well as their predictors has not yet been investigated in CTO patients. A large retrospective registry was used including all consecutive patients with ventricular tachyarrhythmias undergoing coronary angiography at index from 2002 to 2016. Only ICD recipients with CTO were compared to patients without (non-CTO). Kaplan-Meier and Cox regression analyses were applied for the primary end point of first recurrence of ventricular tachyarrhythmias at 5years. Secondary end points comprised of the different types of recurrences, first appropriate ICD therapy and all-cause mortality at 5years. From a total of 422 consecutive ICD recipients with ventricular tachyarrhythmias at index, at least one CTO was present in 25%. CTO was associated with the primary end point of first recurrence of ventricular tachyarrhythmias at 5 years (55% vs. 39%; log rank p = 0.001; HR = 1.665; 95% CI 1.221-2.271; p = 0.001), as well as increased risk of first appropriate ICD therapy (40% vs. 31%; log rank p = 0.039; HR = 1.454; 95% CI 1.016-2.079; p = 0.041) and all-cause mortality at 5years (26% vs. 16%; log rank p = 0.011; HR = 1.797; 95% CI 1.133-2.850; p = 0.013). Less developed collaterals (i.e., either ipsi- or contralateral compared to bilateral) and a J-CTO score ≥ 3 were strongest predictors of recurrences in CTO patients at 5years. A coronary CTO even in the presence of less developed collaterals and more complex CTO category is associated with increasing risk of recurrent ventricular tachyarrhythmias at 5years in consecutive ICD recipients.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call