Abstract

The current paper represents a round loop for calibration of torque tools organized by NIS. These measurements are aimed to evaluate the performance of torque tools calibration laboratories in Egypt in the range from 200 N m up to 1000 N m in both clockwise and anticlockwise directions. Results are evaluated using En-value method and Z-score method based on the weighted means and the corresponding uncertainties. The outlier test and consistency test were performed on the measurements before applying the Z-score test. The Z-score evaluation shows that participants’ results are consistent specifically in the range from 400 N m up to 1000 N m. Further comparison with lower capacity artifact is recommended in the lower torque range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.