Abstract

A negative oxygen ion source is under development to produce O- and O2 - beams used for a secondary ion mass spectrometer at Institute of Modern Physics (IMP), and both filament and radio frequency (RF) schemes are tested. The filament driven ion source, which was initially designed for H- production, has a 10-pole multicusp plasma chamber, two sets of virtual magnetic filters, and a 3-electrode extraction system. The RF scheme, which is improved by changing a RF back plate from the filament ion source, has an external planar spiral RF antenna behind an AlN window. The RF power system consists of a continuous wave (CW) 13.56 MHz/2 kW power supply, a capacitive automatching network and a water-cooled flat RF antenna made from a 6-mm copper tube. Oxygen and carbon dioxide gases are used to produce O- and O2 - ion beams, and ion composition is analyzed by using a Wien filter. However, the lifetime of filaments sustains from several minutes to hours before fracture occurred while the RF one can work stably with no maintenance, and the experiment results will be given on O- ion production with two different technologies using our ion source test facility.