Abstract

Two new generation carbon dioxide absorbents, DrägerSorb Free and Amsorb Plus, were studied in vitro for formation of compound A or carbon monoxide, during minimal gas flow (500 ml x min(-1)) with sevoflurane or desflurane. Compound A was assessed by gas chromatography/mass spectrometry and carbon monoxide with continuous infrared spectrometry. Fresh and dehydrated absorbents were studied. Mean (SD) time till exhaustion (inspiratory carbon dioxide concentration >or= 1 kPa) with fresh absorbents was longer with DrägerSorb Free (1233 (55) min) than with Amsorb Plus (1025 (55) min; p < 0.01). For both absorbents, values of compound A were < 1 ppm and therefore below clinically significant levels, but were up to 0.25 ppm higher with DrägerSorb Free than with Amsorb Plus. Using dehydrated absorbents, values of compound A were about 50% lower than with fresh absorbents and were identical for DrägerSorb Free and Amsorb Plus. With dehydrated absorbents, no detectable carbon monoxide was found with desflurane.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.