Abstract

One of the critical issues while operating bolometric detectors over periods of time of 1 year or more consists of keeping their response stable within a 0.1% level, despite the unavoidable temperature fluctuations of the cryogenic set-up. By using an energy pulser, which periodically delivers a fixed amount of energy in the absorber, it is possible to stabilize the response of the bolometers. A stabilization technique using heating devices, made up of heavily doped semiconductor material (well above the metal-to-insulator transition), has been developed in the framework of the CUORE experiment. In this paper we describe in detail the procedure for the realization of the heating elements, based on silicon planar technology. We then report on the multi-step low temperature characterization (77 K, 4.2 K, 1.5 K, 35 mK) of the heaters. Finally, an example of achieved stabilization for a CUORE-like detector is reported. The ∼ 1500 heaters tested at ∼ 1.5 K show less than 0.5% change in resistance between 30 μ V and 3 mV, and less than 1% change in value between 50 mK and 800 mK. In particular, the resistance change between 4.2 K and 1.5 K is less than 0.1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.