Abstract

Ultrasound signals reflected from the human liver contain significant information about its condition. Several techniques have been developed to process such signals. In this paper, two approaches are described for estimating the slope of the acoustic attenuation coefficient, denoted by β, which provides important cues about the condition of diffuse liver disease. The first is a nonparametric approach which determines the β value from the slope of log periodogram differences. The second is a parametric approach which requires a Gaussian-shaped spectrum and determines the β value from s shift in the spectral center-frequency. A zero-crossing count procedure allows the parametric approach to be implemented with minimal hardware and in real-time. Simulated and actual signals illustrate the two approaches.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.