Abstract Mass spectrometry has become a well-established analytical method in the forensic identification of drugs and explosives [1,2]. In electron impact (EI) mass spectrometry (MS) the investigated sample is ionized by an electron beam having an energy of 70 eV. Because of this high ionization energy, the ions, after their formation, will usually decompose to fragmentation products including both charged and neutral species. The complexity of the EI mass spectrum, while often an important asset because of its “fingerprint” value, is clearly a disadvantage when complex mixtures are to be analyzed. When the compound is more complex, the amount of fragment ions is larger and the chances of observing a molecular ion are smaller. Electron impact mass spectrometry can be applied to mixtures, but only after separation, usually performed by a gas chromatograph-MS combination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.