Abstract

This is the third of a series of articles detailing the development of near-infrared spectroscopy methods for solid dosage form analysis. Experiments were conducted at the Duquesne University Center for Pharmaceutical Technology to develop a system for continuous calibration monitoring and formulate an appropriate strategy for calibration transfer. Indicators of high-flux noise (noise factor level) and wavelength uncertainty were developed. These measurements, in combination with Hotelling's T(2) and Q residual, are used to continuously monitor instrument performance and model relevance. Four calibration transfer techniques were compared. Three established techniques, finite impulse response filtering, generalized least squares weighting, and piecewise direct standardization were evaluated. A fourth technique, baseline subtraction, was the most effective for calibration transfer. Using as few as 15 transfer samples, predictive capability of the analytical method was maintained across multiple instruments and major instrument maintenance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.