Abstract

Previous case series have described the safety and efficacy of different stent models for stent-assisted aneurysm coiling (SAC), but comparative analyses of procedural results are limited. This study investigates the procedural outcome and safety of three different stent models (Atlas™, LEO+™ (Baby) and Enterprise™) in the setting of elective SAC treated at a tertiary neuro-endovascular center. We retrospectively reviewed all consecutively treated patients that received endovascular SAC for intracranial aneurysms between 1 July 2013 and 31 March 2020, excluding all emergency angiographies for acute subarachnoid hemorrhage. The primary procedural outcome was the occlusion rate evaluated with the Raymond–Roy occlusion classification (RROC) assessed on digital subtraction angiography (DSA) at 6- and 12-month follow-up. Safety assessment included periprocedural adverse events (i.e., symptomatic ischemic complications, symptomatic intracerebral hemorrhage, iatrogenic perforation, dissection, or aneurysm rupture and in-stent thrombosis) and in-house mortality. Uni- and multivariable logistic regression analyses were performed to identify patient baseline and aneurysm characteristics that were associated with complete aneurysm obliteration at follow-up. A total of 156 patients undergoing endovascular treatment via SAC met the inclusion criteria. The median age was 62 years (IQR, 55–71), and 73.7% (115) of patients were female. At first follow-up (6-month) and last available follow-up (12 and 18 months), complete aneurysm occlusion was observed in 78.3% (90) and 76.9% (102) of patients, respectively. There were no differences regarding the occlusion rates stratified by stent model. Multivariable logistic analysis revealed increasing dome/neck ratio (adjusted odds ratio (aOR), 0.26.; 95% CI, 0.11–0.64; p = 0.003), increasing neck size (aOR, 0.70; 95% CI, 0.51–0.96; p = 0.027), and female sex (aOR, 4.37; 95% CI, 1.68–11.36; p = 0.002) as independently associated with treatment success. This study showed comparable rates of complete long-term aneurysm obliteration and safety following SAC for intracranial aneurysm with three different stent-models highlighting the procedural feasibility of this treatment strategy with currently available stent-models. Increased neck size and a higher dome/neck ratio were independent variables associated with less frequent complete aneurysm obliteration.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call