Abstract

The aim of this study was to investigate why capnographs malfunction at high altitude, in order to suggest ways of improving instrument reliability. Four machines were decompressed to simulated high altitude and measurements taken. The study showed that the reduced air density at altitude has effects upon the pump, causing difficulty in entraining gas, and that high altitude has additional effects upon calibration of the instruments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.