Implantable cardioverter-defibrillators (ICDs) have been a successful adjunct to the management of arrhythmias in patients with Long QT syndrome (LQTS). In two patients, interactions between LQTS and the ICD were diagnosed and corrected. Oversensing of T waves was confirmed in the first, while in the second, the arrhythmia disappeared when T wave abnormalities improved after cessation of H2 blocker therapy. In patients with LQTS and an ICD, T wave oversensing should be considered. Interventions that may have an adverse effect on repolarization should be avoided.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.