Abstract

Dual-sided subdural electrodes are used in the localization and lateralization of seizure-onset zones when the area of interest is within the interhemispheric fissure. We designed the current study to test the validity of the assumption that each side of the dual-sided electrodes records from the hemisphere it faces. We recorded with dual-sided strip and grid electrodes implanted in the occipital interhemispheric space in two patients with nonoccipital epilepsy during two visual stimulation tasks in which subjects were presented with visual stimuli in the ipsilateral or contralateral visual hemifields. Our findings show substantial contamination of recordings from the opposite hemisphere. Although, as expected, electrodes recording through the falx record faintly from the contralateral cortical surface, they unexpectedly pick up strong signals from the cortex behind them. Therefore, we conclude that these electrodes should not be used for lateralization of the origin of epileptic activity or evoked responses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.