Abstract
Abstract Detectors based on TlBr crystals are ideally suited for the probes, where small size and rapid stopping power are highly desirable attributes. The development results of TlBr probes for medical and nuclear industry applications are presented here. Probes for both applications include TlBr detectors with dimensions of 5×5×2 mm 3 and a microelectronics preamplifier. The different construction and technological aspects of probes designed for both application types are discussed. The probes were tested over a wide energy range. We obtained energy resolutions of 3.6, 4.9 and 14.5 keV at energies of 59.6, 122 and 662 keV, respectively, for the developed probe prototypes at room temperature. The observed performances of probe prototypes having TlBr detectors confirmed their promising potential for probe applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.