Abstract
To enhance the qualitative information gained by photographing the optic nerve head, we built a silicon photovoltaic detector (photodiode) in the image plane of a fundus camera (Zeiss). It permitted quantitative evaluation of the relative brightness of the illuminated papilla. Following construction of the device and during the early course of more than 260 measurement sessions (on more than 130 subjects), several heretofore unknown sources of error were encountered and subsequently dealt with, resulting in a margin of error for the entire system of between 0.3% and 1.2%. Pallor of the optic nerve head induced by artificial ocular hypertension can be well reproduced by this method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.