Abstract

Proper left ventricular assist device (LVAD) insertion will help maximize LVAD flow and may reduce adverse events such as right heart failure and pump thrombosis. Although no standardized insertion technique has been universally accepted, the goals are: unobstructed inflow cannula, unobstructed outflow graft with avoidance of right ventricular compression, and prevention of pump migration. To achieve these objectives for the HeartMate II LVAD, we delineate four principles: proper pump pocket creation, optimized positioning of inflow cannula and outflow graft, proper pump position in the body, and fixation. These basic principles are easy to implement and have been beneficial in our patients, assuring long-term unobstructed LVAD flow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.