Abstract

A novel, simple and sensitive method based on the use of a PRiME (process, robustness, improvements, matrix effects, ease of use) pass-through cleanup procedure combined with ultra-fast liquid chromatography-tandem quadrupole mass spectrometry (UFLC-MS/MS) for the determination of 4 AFs, i.e., AFB1, AFB2, AFG1 and AFG2 in human serum, was developed and validated. The effective PRiME pass-through cleanup procedure using the Oasis PRiME HLB cartridge was optimized for the extraction of AFB1, AFB2, AFG1 and AFG2 in human serum. The obtained results demonstrated the higher extraction capacity of the PRiME pass-through cleanup procedure with recoveries between 92.1% and 102%. The limits of detection (LODs) for AFB1, AFB2, AFG1 and AFG2 were 0.048 μg L−1, 0.069 μg L−1, 0.043 μg L−1 and 0.14 μg L−1, respectively. Validation results on linearity, specificity, accuracy and precision, as well as on application to the analysis of AFs in 10 liver cancer volunteers demonstrated the applicability to clinical studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.