Abstract

To examine the management and possible causes of primary valve malfunction of the Krupin eye valve with disk. The authors reviewed the results of 113 patients undergoing implantation of the Krupin eye valve with disk and identified eight patients with primary valve malfunction requiring surgical revision. Valve revision involved manipulation (n = 1 case), explantation of the malfunctioning valve and implantation of a new valve (n = 2), and amputation of the valve (n = 5). Six of eight patients had final intraocular pressures of < 21 mmHg on one or no medications at a mean interval of 15.9 months (range 5-36) after surgical revision. Transient postoperative hypotony was noted in three patients and chronic hypotony with loss of light perception in one patient. One explanted valve was examined and found to have partially fused leaflets. Surgical revision in cases of primary valve malfunction of the Krupin eye valve with disk may be accomplished relatively safely with an acceptable level of postoperative complications. The etiology of primary valve malfunction may be related to the sterilization process and prolonged storage before implantation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.