Abstract

The paper presents a technique to detect significant systolic peaks, the percussion (P) and tidal peak (T) and diastolic peak (D) from the arterial blood pressure (ABP) waveform. The technique is aimed at robust detection even in presence of significant noise. Singular Value Decomposition (SVD) based dominant period extraction of the ABP waveform followed by wavelet transform and local peak detection is applied to detect the points of interest. MIMIC-II ABP databse serves as a training dataset to select SVD and wavelet transform parameters and CSL Benchmark database is used to analyze the technique. Salient systolic peak detection for the CSL dataset was performed with positive predictive value and sensitivity figures of 98.48% and 99.24% respectively.