Abstract

The 4πβ-γ coincidence counting and CIEMAT/NIST liquid scintillation method were used in the standardization of a solution of 152Eu. In CIEMAT/NIST method, measurements were performed in a Liquid Scintillation Counter model Wallac 1414. In the 4πβ-γ coincidence counting, the solution was standardized using a coincidence method with ‘‘beta-efficiency extrapolation”. A simple 4πβ-γ coincidence system was used, with acrylic scintillation cell coupled to two coincident photomultipliers at 180° each other and NaI(Tl) detector. The activity concentrations obtained were 156.934 ± 0.722 and 157.403 ± 0.113 kBq/g, respectively, for CIEMAT/NIST and 4πβ-γ coincidence counting measurement methods.