Abstract

A solution of 225Ac was standardized by NRC using the triple-to-double coincidence ratio (TDCR) method. The counting efficiencies were calculated assuming a counting efficiency of 100% for alpha decays and those calculated using the MICELLE2 Monte Carlo code for beta decays and was approximately 500% for the NRC TDCR system. The relative uncertainty for the activity concentration was determined to be 0.25%. This agreed with measurements performed using gamma spectroscopy and a predicted calibration factor for the Vinten 671 ionization chamber as calculated using an EGSnrc model, implementing radioactive decay. Finally, the half-life of 225Ac was determined from long-term measurements using ionization chambers and liquid scintillation counting. The NRC measured half-life for 225Ac was found to be 9.914(4) days and is consistent within an expanded uncertainty coverage of k = 2 with the most recent (Kossert et al., 2020; Pommé et al., 2012) measurements of this decay parameter.