Abstract

High loosening rates of the trapezium components of trapeziometacarpal total joint prostheses have been reported. The purpose of this study was to compare the primary press fit fixation of two different, uncemented cup designs (MOTEC and Elektra) with the primary fixation of a cemented polyethylene cup (DLC) in a pig bone model. We did a push out test to measure the maximal load strength of the implants and a low-pressure cyclic loading test combined with radiostereometric analysis to measure the micromotion of the implants. There was no significant difference in fixation strength between the two uncemented screw cup designs or between the two uncemented screw cups and the cemented polyethylene cup. However, we found that threading of the bone before insertion of the Electra screw cup weakens the primary fixation strength of the implant. The results indicate that focus should be on the insertion technique as well as on the cup design of uncemented trapezium cup implants. Further studies of trapezium implant migration in a clinical setting are needed.