Abstract

This paper presents a system developed at the Vibration Laboratory of Inmetro, which is used for primary calibration of vibration transducers by the fringe counting method. This system includes a vibration exciter, a Michelson interferometer, a data acquisition board, a band-pass filter, a universal counter and a software for measurement automation. It allows the laboratory to perform calibrations in accordance with the international standard ISO 16063-11 in the frequency range from 0.4 Hz to 160 Hz. Some experimental results are presented herein.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.