Abstract

The activity concentration of a solution of 99Tc was absolutely determined by liquid scintillation measurements based on the triple-to double-coincidence ratio method (TDCR) and compared with the results given by 4πβ(LS)-γ(NaI) live-timed anticoincidence (with extending dead-time) and classical 4πβ(PC)-γ(NaI) coincidence counting systems based on using 60Co as the efficiency-tracing radionuclide. The results of anticoincidence and coincidence counting are, respectively, 0.18% and 0.66% higher than the result from TDCR measurements, but they are consistent within uncertainties.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call