Abstract
Primary activity standardization of 177Lu has been conducted at the Center for Technology of Radiation Safety and Metrology-National Nuclear Energy Agency (PTKMR-BATAN) to establish a national reference standard for 177Lu measurements. The standardization was performed using a 47i(3(LS)-y coincidence counting system with digital coincidence counting (DCC) technique. The activity standardization of 177Lu gave a result of 148 kBq/g + 1.78 kBq/g at a reference time with uncertainty value evaluated at k = 2. The result was then used to establish a secondary standard by calibrating the reference ionization chamber using the ‘dialing-in’ method. A dial setting value of 024 was obtained on the standard reference IC. The new dial setting will be used in the routine calibration services of commercial dose calibrator to ensure all measurements of 177Lu are traceable to the national standards.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.