Abstract

The radioactive concentrations of (166m)Ho, (134)Cs and (133)Ba solutions have been standardised using a 4πβ-4πγ coincidence counting system we have recently set up. The detection in the beta channel is performed using various geometries of a UPS-89 plastic scintillator optically coupled to a selected low-noise 1in. diameter photomultiplier tube. The light-tight thin capsule that encloses this beta detector is housed within the well of a 5in.×5in. NaI(Tl) monocrystal detector. The beta detection efficiency can be varied either by optical filtering or electronic discrimination when the electrons loose all their energy in the plastic scintillator. This 4πβ-4πγ coincidence system improves on our 4πβ(PC)-γ system in that its sample preparation is less labour intensive, it yields larger beta- and gamma-counting efficiencies thus enabling the standardisation of low activity sources with good statistics in reasonable time, and it makes standardising short-lived radionuclides easier. The resulting radioactive concentrations of (166m)Ho, (134)Cs and (133)Ba are found to agree with those measured with other primary measurement methods thus validating our 4πβ-4πγ coincidence counting system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.