Abstract

The risk of infection transmitted during mouth-to-mouth or mouth-to-nose resuscitation procedures is difficult to define but is possibly quite low. However, the perceived risk is sufficient to cause serious concern for many individuals, including trained hospital personnel as well as the general public, and may preclude prompt and effective action. A novel airway device was evaluated for the retention of infective droplets and fluid permeability under simulated resuscitation conditions using a cardiopulmonary resuscitation training manikin. Retention of a 0.5-5.0 micron aerosol of Staphylococcus aureus cells was greater than 80% at flow rates of 6 l/min while under simulated resuscitation conditions the trapping of bacteria, originating predominantly from saliva, was over 90%. These data suggest that this device may afford significant protection against transmission of infection during exhaled air resuscitation manoeuvres.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.