Abstract

Patients with implantable cardioverter defibrillator lead insulation failures may present with oversensing and/or abnormal impedance. The Lead Integrity Alert (LIA) monitors right ventricular pace/sense leads using both continuous oversensing and daily impedance measurementd. Oversensing consists of isolated short R-R intervals and nonsustained runs of short R-R intervals. The LIA algorithm has been studied for Sprint Fidelis conductor fractures, but not for lead insulation failures. We report on a patient with a failed St. Jude Riata™ ST lead (St. Jude Medical, St. Paul, MN, USA) connected to a Medtronic Virtuoso DR (Medtronic Inc., Minneapolis, MN, USA) with the LIA. Oversensing triggered the LIA, while the impedance trend was normal.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.