Abstract
To assess the prevalence of carotid stent fractures and quantify any associated restenosis or clinical events in a single-center experience. Seventy-eight consecutive patients were recorded in a prospective database between 2003 and 2007. Nearly two thirds (51, 65%) were available for follow-up, which included magnified multiplanar radiography and selective duplex ultrasonographic studies to evaluate fracture and restenosis in 53 self-expanding stents (49 open-cell and 4 closed-cell designs). The images were assessed independently by 2 vascular radiologists blinded to the patients' clinical data. There was 1 (1.9%) stent fracture with no associated adverse events or restenosis. Six (11%) stents had an irregular "fish-scale" appearance that was easily confused for fracture and seen only in the open-cell stents. Stent fractures do occur in the carotid artery; however, they are rare and, in our experience, have no clinical sequelae. Radiological assessment should be incorporated into clinical and ultrasound surveillance protocols.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.