Abstract

A new method for sensitive and fast screening of melamine (MEL) in milk products was developed with a low-cost disposable microfluidic device coupled with ultraviolet (UV) detection. This method avoided the need of sample pretreatment prior to the separation process, thus was simple and green. Due to the advantages of the device and fracture sampling technique, milk sample could be directly sampled through the fracture to achieve baseline separation from amino acids, and proteins in the sample did not interfere with the detection. Using 20 mM phosphate running buffer (pH 9.0), a sampling time of 3 s at +180 V and a separation voltage of +1800 V (240 V/cm), this method could detect MEL in milk within 75 s. At the detection wavelength of 202 nm, the linear range for MEL was from 1.0 to 100 μg mL −1 with a detection limit of 0.23 μg mL −1 ( S/ N = 3). The novel protocol had been successfully used to screen MEL in milk samples with recovery more than 82%. The environmentally friendly methodology for pretreatment-free sensitive screening of MEL provided promising applications in monitoring the quality of different foods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.